CCR1000 Specifications

  • Temperature range of ambient to 1000°C

  • Heating rate up to 200°C/min

  • Temperature stability <1°C

  • Pressure 5Bar max

  • Maximum sample volume 164.03mm3

  • Sample size 5.9mm Ø, 6mm Ø internal height, accommodates between 70-100mg (dependent on sample type)

  • Samples are loaded on low cost disposable ceramic fabric filters

  • Standard size 1/16 gas analyser high pressure connectors

  • Compatible with reflected light upright microscopes and Raman

  • Minimum objective working distance 8.0mm

  • Easily replaced quartz transparent window

  • Simple removable ceramic sample holder

  • All critical component parts manufactured from non reactive materials

  • Water cooling to keep stage body at safe temperature

  • Gases remain at elevated temperatures preventing condensation before analysis

  • Fits 160mm x 110mm XY table recesses but can be adapted to fit motorised XY tables with alternative recesses